prepFAST for pharmaceutical laboratories
Automated elemental impurity analysis for USP <232> & <233>
The prepFAST inline autodilution system, paired with ICP-OES or ICP-MS, fully automates the USP <233> method — from calibration through spike recovery — with no manual dilutions.
What prepFAST automates
- Dilutes a single stock standard to build linear calibration curves.
- Dilutes samples to the appropriate total dissolved solids.
- Spikes samples at user-defined J values for QC and validation.
Select any figure below to enlarge it.
About USP <232> and <233>
USP Chapter <232> defines a target (J) value as a function of the permitted daily exposure (PDE, µg/day) and a drug’s maximum daily dose (g/day).
Chapter <233> sets out the procedures for determining elemental impurities by ICP-OES and ICP-MS. It requires a calibration curve and QC validation — spike recovery, accuracy, precision and stability — all based on the target J value.
Step 2
Automated sample dilution
Inline dilution to the right total dissolved solids keeps the plasma stable and surpasses USP validation criteria.
Step 3
Auto “J” spiking
prepFAST spikes QC and validation samples at user-defined J values automatically, so the analyst prepares no spikes by hand.
- Spike recovery of 92–108% for all elements by ICP-OES.
- Well inside the USP <233> requirement of 70–150%.